Screening Plasma for IgA Deficiency in Under an Hour
A label-free sandwich assay on the P4SPR gives a deficient / normal answer from plasma in about 32 minutes. By ELISA, the same test takes a day.
The need
IgA deficiency is the most common immune deficiency: about 0.2% of Canadian blood donors, roughly 1 in 600. Patients who have it can react severely to a transfusion of normal blood, so blood services keep IgA-deficient donors on a registry and hospitals sometimes need to test a patient before transfusing. Today that test is an ELISA run by a central laboratory, with a result after 24 hours or more.
The question for this work: can the same test run at the hospital or the blood drive, within an hour, without a central lab?
How it works on the P4SPR
The assay is an ELISA-style sandwich, read without labels. An anti-IgA antibody on the gold sensor captures IgA from the diluted plasma, and a second anti-IgA antibody then binds to the captured IgA. SPR reads the mass of that second antibody directly, so there is no enzyme, substrate or colour step. Plasma proteins that stick to the surface don't count, because the reading comes from the second antibody only. The gold sensor carries AffiCoat, our low-fouling coating, which keeps most plasma proteins off the surface in the first place.
One run takes five steps on the instrument (Figure 1). The four channels hold two replicates of the sample, a positive reference and a negative reference, so every result comes with its own controls.
What the study showed
| Result | |
|---|---|
| Time to result | About 32 min once plasma is separated; under 1 h in total (ELISA: 24 h or more) |
| Agreement with ELISA | Same deficient / normal call on every clinical plasma sample tested |
| Sensitivity in plasma | Limit of detection below the 500 ng/mL clinical cut-off |
Full method and data: Dubois et al., Transfusion 2024 (reference below).
Why the P4SPR fits this test
| What a clinical lab needs | What the P4SPR brings |
|---|---|
| A result the same day | Real-time readout: about 32 minutes per sample, no overnight incubation |
| Testing outside a central lab | Portable, bench-top, USB-powered; no moving parts |
| Confidence in each result | Four channels: sample in duplicate plus positive and negative references in the same run |
| Works in real samples | AffiCoat low-fouling surface and a second-antibody readout that ignores plasma background |
| Low cost per test | No labelled antibody or substrate; a sensor can be regenerated and reused for several samples |
| An assay your team can build | Developed like an ELISA: same antibodies, same logic, no labelling chemistry |
Beyond IgA
The design is not specific to IgA. Swap the antibody pair and the same capture-and-confirm format applies to other proteins in blood. As a proof of concept, the same format was set up for ferritin, the iron-storage protein checked in frequent blood donors. Reading the second antibody gave about 2.5 times more signal than reading the captured ferritin directly, and a detection limit about half as low (Figure 2).
If you run an ELISA today and need the answer faster, or closer to the patient, the method in this note is a starting point.
What you need
| Item | Source |
|---|---|
| P4SPR instrument | Affinité Instruments |
| Gold sensors with AffiCoat | Affinité Instruments |
| Antibody immobilization (EDC/NHS, ethanolamine) | Standard reagents; see our covalent immobilization protocol |
| Anti-IgA antibody (capture and detection) | Commercial polyclonal anti-human IgA |
| Positive and negative reference plasma | Normal plasma and IgA-deficient plasma |