←  Resources  /  Technical notes

Tech noteTN-09

Getting a clean signal

TopicReferencing, buffer match, drift
Reading time6 min

IN SHORT

  • Four parts: every raw trace mixes real binding, a jump from the sample liquid itself (bulk effect), unwanted sticking, and slow drift.
  • Double referencing: subtract the reference channel, read within the same ~1 s cycle (TN-03: Lensless optics and wavelength readout), then a buffer-only run. What is left is binding.
  • Before you start: match every sample to the running buffer, and give a new chip time to settle.

Overview

The trace from a sample channel is never binding alone.

This note takes the raw signal apart, shows how referencing removes what you don’t want, and covers the two hardest cases: buffer mismatch and complex samples.

1 · WHAT’S IN A RAW TRACE

What’s in a raw trace

Every sample channel records the sum of four things: the specific binding you want, and three things you don’t.

Figure 1 — signal parts
Fig. 1. Left: the four parts of a sample-channel signal: specific binding, a bulk step while the sample is in, slow drift, and non-specific binding that does not wash off. Right: their sum, which is what the channel records. Simulated.
ComponentWhat causes itHow to remove it
Specific bindingYour analyte binding your ligandThat’s the signal you keep
Bulk effectThe sample’s refractive index differs from the running bufferReference channel, buffer matching
DriftTemperature, the surface settling, slow loss of captured ligandEquilibration, blanks
Non-specific bindingOther molecules sticking to the surfaceBlocking, low-fouling surfaces, a reference surface

2 · REFERENCING

Referencing, in two steps

Step 1, the reference channel. One channel carries a surface without your ligand, or with an irrelevant protein, and sees the same sample. Subtracting it removes the bulk effect, shared drift and much of the non-specific binding. On our instruments every channel shares the same optical block and geometry, so the reference sees exactly what the samples see (TN-03: Lensless optics and wavelength readout).

Step 2, buffer blanks. Inject running buffer as if it were a sample, a few times through the run. Subtracting these blank injections, after the reference, removes what’s left of the injection artefacts and systematic drift. Together the two steps are called double referencing.

Figure 2 — double referencing
Fig. 2. Double referencing as arithmetic. (1) The sample channel and the reference channel, read in the same cycle, both carry the bulk step and drift. (2) Subtracting the reference removes most of it; a buffer blank, processed the same way, shows what is left. (3) Subtracting the blank leaves the binding. Simulated. The clean curve is then fitted (TN-10: From sensorgram to answer).

3 · THE BULK EFFECT

Match your buffers

A small difference in buffer composition can produce a larger signal than real binding, especially for small molecules (TN-07: Designing a binding experiment).

DMSO. Compounds dissolved in DMSO shift the signal strongly if the sample and running buffer contain different amounts. Keep the DMSO concentration identical in every sample and in the running buffer, and run a short DMSO calibration series if it varies.

Glycerol and salt. Protein storage buffers often contain glycerol. Carried into the sample, it shows up in every channel at once, which can look like binding. Our protein–protein screening note shows exactly this, solved by buffer exchange.

The rule. Dilute samples into running buffer, or exchange them into it.

4 · NON-SPECIFIC BINDING

Non-specific binding

Block unused reactive groups after coupling (TN-06: The binding surface).

Use a low-fouling surface when the sample is crude: peptide monolayers, such as AffiCoat, strongly reduce protein adsorption from serum.

Use a matched reference surface: the same chemistry, with an irrelevant protein in place of your ligand, catches what sticks to “a protein-coated surface”, not just to bare chemistry.

Figure 3 — serum fouling
Fig. 3. Serum over a standard surface and over a low-fouling peptide surface. Both show the same bulk step; on the standard surface proteins keep accumulating and stay after the buffer wash. Illustrative shapes, not measured data.

5 · DRIFT

Drift

After inserting a chip, or changing buffer, the baseline drifts while temperature and the surface settle. Start measuring once the baseline is flat. A drift test on a new chip, buffer only, tells you how long that takes on your setup.

6 · COMPLEX SAMPLES

Complex samples

Serum, plasma, lysate and culture media are full of proteins that can stick, and their refractive index differs from buffer.

Dilute where the assay allows, and match the running buffer to the diluted matrix. Use low-fouling surfaces and a matched reference (sections 4 and 2).

Remove what you can before the sensor. For whole blood, a plasma-separation chip placed in front of a P4SPR let biomarkers be measured from blood.

Not ready to buy?

Send us your samples or book mobile SPR with Affibio: we run your molecules on P4SPR 2.0 and share the data.